Developments in near and mid infrared fibre optic materials have enabled Fibre Photonics to provide a family of ATR immersion probes. Reliable spectral measurements of fundamental molecular vibrations in the ‘fingerprint region’ are now a reality for liquid chemical processes in the near and the mid infrared regions in applications such as:
|
Reaction Monitoring Example
|
|
ATR Tip Material |
Fibre Optic Material |
Transmission |
Temperature |
Pressure* | |
|
(cm-1) |
(mm) |
(°C) |
(Bar) | ||
|
Diamond |
Polycrystalline IR |
600 – 1900 2300 – 3100 |
5.2 – 17 3.2 – 4.5 |
-150°C / +140 |
200 |
|
Zinc Selenide |
Polycrystalline IR |
600 – 3100 |
3.2 – 17 |
-150°C / +140 |
10 |
|
Cubic Zirconium |
Chalcogenide IR |
1550 – 6650 |
1.5 – 6.5 |
-150°C / +90 |
100 |
* For 12mm diameter shafts
|
Common ATR Probe Specifications | |
|
Probe Head (Shaft) Material |
Hastelloy C22 |
|
Sheating Material |
Liquid Tight SS-Conduit, KOPEX |
|
Minimal Bending Radius (mm) |
130 |
|
Probe Body/Shaft Diameter (mm) |
6.3 or 12 |
|
Probe Body Length (mm) |
200, 250, 300, 350, 400, 450, or 500 * |
|
Total Length (Body plus Fibre) (m) |
1.5, 2.0, 2.5, 3.0, 3.5, 4.0, 4.5 or 5.0 * |
|
Input/ Output connectors for legs |
Long Ti-SMA (opt.: any other type) |
|
Compatible Process-Interfaces |
Ceramat-FOS or SensoGate-FOS ** |
* Customized dimensions are available on request
** Available for 12mm shafts only
The Fibre Photonics range of ATR immersion probes are manufactured from a range of corrosion and abrasion resistant materials. They are suitable for laboratory, pilot plant and in-process applications and are compatible with all types of spectrometers.User selectable tip and shaft materials allow each probe to be customised to meet the needs of applications, which include:



Click here for More Examples
For product availability and pricing contact: sales@fibrephotonics.com +44 (0) 845 557 6450