Universal, pneumatically operated retractable fitting with modular design
High Availability
The sensor lock-gate principle allows rinsing the O-rings during the stroke movement. Therefore, the integrated O-rings are notably less exposed to stress than in conventional fittings. Wear is reduced and durability is increased.
Safety
The point of separation between the stroke cylinder and the calibrating chamber is double sealed and ventilated. This prevents media from escaping from the calibrating chamber into the drive unit, and conversely prevents compressed air from entering the calibration chamber. SensoLock prevents the immersion in the process when the sensor has been removed.
Plug & Play Installation
The central multiplug contains tubings for rinsing, cleaning, and calibration media, including limit switches and check valves and enables a fast, safe, and space-saving connection.
Easy Maintenance
The particular ease of maintenance of the sensor lock-gate is achieved by a well thought-out design that allows performing virtually all maintenance on the site without the aid of a workshop. Drive and immersion tube can simply be taken off by loosening a coupling nut, without complete dismantling. The immersion tube is divided by a bayonet coupling to allow replacing only the particularly stressed process-wetted part including calibration chamber which helps to reduce costs.
Efficient Rinsing
Another unique feature is the tangential routing of the rinsing media with high flow rates for optimum cleaning effect.
The Facts:Specifications:
| Ambient temperature | 5 ... 70 °C |
| Permitted process temperature |
1.4571 stainless steel, Hastelloy: 5 ... 140 °C |
|
Permitted process pressure during movement |
6 bars gauge pressure (1.4571, Hastelloy: 140 °C) |
|
Permitted static pressure (service position) |
16 bars (40 °C) |
| Air supply |
4 ... 7 bars (filtered 10 ... 25 ľm, oil- and condensate-free) |
| Effective immersion depth of the sensor | 65 mm or 120 mm, measured from lower edge of flange to sensor tip (flange version) |
|
Total immersion depth incl. immersion tube |
88 mm or 143 mm, measured from lower edge of flange to immersion tube tip (flange version) |
| Process adaptations |
Flanges (lapped) DN 32 / DN 50 / DN 65 / DN 80 / DN 100 |
| pH Sensor options |
225 mm, Pg 13.5 |
| Rinse medium inlet |
1) for UnicalŽ 9000 / UnicleanŽ 900: multiplug connection |
| Rinse medium outlet |
Connection for hose grommet with coupling |
| Rinse function | Seal rinsing during stoke movement |
| Process-wetted materials |
1.4571 stainless steel, Hastelloy |
| Housing |
1.4571 stainless steel |
| Protection class |
IP 66 |
|
Connection to UnicleanŽ 900 / UnicalŽ 9000 |
Compact multiplug connection |
| Free connection |
1 x 4/6 mm hose coupling for rinse water/cleaning agent, |
| Sensor protection in process |
Integrated in immersion tube |
| Limit switches |
Non-contacting electrical limit position control in conjunction with UnicleanŽ 900 / UnicalŽ 9000 / ZU 0631 |
| Certificates |
3.1 Material Certificate (optional) ATEX II 1GD cII |